Main Body |
High-strength engineering plastic rack, light, beautiful and corrosion resistant |
Scope of application |
Adult, child |
Gas Source |
O2, N2O:0.28-0.6MPa N2O: 0.28~0.6MPa |
Flow Meter |
O2: 0.05-10L/minN2O: 0.05~10L/min |
O2, N2Olinkage and N2O Stopper |
When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. |
Flow Rate of Rapid Oxygen Supply |
25-75L/min |
Low Oxygen Pressure Alarm |
There will be sound alarm when the oxygen pressure< 0.2 MPa |
Vaporizer |
It has the function of automatic compensation based on pressure, temperature and flow rate. The regulation range of evaporator concentration is 0-5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.And vaporizers that are imported with original packaging are also available. |
Low oxygen pressure alarm |
There will be sound alarm when the oxygen pressure<0.2Mpa |
Respiratory Circuit |
Working Mode: all-close, semi-close, semi-open APL: ≤12.5kPa |
Respiratory Bellows |
Bellows for adults, bellows for children Tidal volume range: 0-1500ml |
Display Mode |
High-definition 5.7'' LCD screen display |
Ventilation Mode |
IPPV, SIPPV, SIMV, MANUAL |
Ventilation Function |
PEEP, SIGH, IRV |
Ventilation Parameters |
Tidal Volume: 50-1500ml Rate: 2-99bpm SIMV: 2-20bpm I:E 2:1-1:8 Inspiratory trigger pressure -10-20cm H2O PEEP 2~20cmH2O Pressure limitation 10~60cmH2O SIGH 1.5 times the inspiratory time/100bpm |
Parameters for Ventilation Monitoring |
Tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure -time waveform, flow -time waveform, PEEP, inspiratory trigger pressure |
Airway Pressure Alarm |
Upperlimit setting range 5~60cmH2O Low limit setting range 0~50 cmH2O |
Minute Ventilation Volume Alarm |
Upperlimit setting range 1~20 L/min Low limit setting range 0~19L/min |
Tidal volume |
Upperlimit setting range 100~1500ml Low limit setting range 0~1400ml |
Sustained high-pressure alarm |
it will give alarm when stress have consistently been higher than 2.5 cmH2O (Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm) |
Power |
AC220V50Hz(UPS, Storage battery) |
*(Mechanical Arms) |
*(End-tidal CO2 monitoring) |
*(Specific interface of waste anesthesia gas) |